§6Worked example
§6.1Worked example — why the reverse-phase figure and the size-exclusion figure disagree
- A long-acting amylin analogue is analysed by both methods. Reverse-phase purity by CEI-MS-01: 98.7 %. Size-exclusion analysis of the same sample: monomer 91.4 %, dimer 6.2 %, higher-order species 2.4 %.
- Both determinations are correct. Under the denaturing acidified organic conditions of reverse-phase chromatography, the non-covalent dimer dissociates and its constituent monomers elute in the main peak. Under the non-denaturing isotonic conditions of size-exclusion chromatography, it survives and is counted separately.
- A certificate reporting only the 98.7 % figure describes a preparation in which 8.6 % of the material is aggregated at physiological pH, and does so without saying anything untrue.
- For a class of molecules whose parent hormone is amyloidogenic and whose entire engineering rationale was to prevent self-assembly, this is not a technicality. It is the quality attribute the compound was designed around, and it is invisible to the assay the channel actually performs.
§6.2Illustrative mass balance
References cited on this page
References are numbered in order of first citation in this document. Each superscript in the text links to its entry below.
- International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use. ICH Q6B Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products. ICH Harmonised Tripartite Guideline 1999;Step 4 version. identifier not held by the Institute
- United States Pharmacopeial Convention. General Chapter ⟨788⟩ Particulate Matter in Injections. United States Pharmacopeia — National Formulary (USP–NF) 2024;USP 2024 Issue 1. identifier not held by the Institute
Identifiers are reproduced only where the Institute holds them. Where a digital object identifier or PubMed identifier is not shown, the Institute has recorded the journal and year and has not constructed an identifier.
Compound Evidence Institute · CEI-MS-11/6 · https://compoundevidence.com/methods/aggregates-sec/worked-example/ · retrieved 30 July 2026